Executive Summary

Informations
Name CVE-2023-25931 First vendor Publication 2023-03-01
Vendor Cve Last vendor Modification 2023-11-07

Security-Database Scoring CVSS v3

Cvss vector : CVSS:3.1/AV:P/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H
Overall CVSS Score 6.8
Base Score 6.8 Environmental Score 6.8
impact SubScore 5.9 Temporal Score 6.8
Exploitabality Sub Score 0.9
 
Attack Vector Physical Attack Complexity Low
Privileges Required None User Interaction None
Scope Unchanged Confidentiality Impact High
Integrity Impact High Availability Impact High
Calculate full CVSS 3.0 Vectors scores

Security-Database Scoring CVSS v2

Cvss vector :
Cvss Base Score N/A Attack Range N/A
Cvss Impact Score N/A Attack Complexity N/A
Cvss Expoit Score N/A Authentication N/A
Calculate full CVSS 2.0 Vectors scores

Detail

Medtronic identified that the Pelvic Health clinician apps, which are installed on the Smart Programmer mobile device, have a password vulnerability that requires a security update to fix. Not updating could potentially result in unauthorized control of the clinician therapy application, which has greater control over therapy parameters than the patient app. Changes still cannot be made outside of the established therapy parameters of the programmer. For unauthorized access to occur, an individual would need physical access to the Smart Programmer.

Original Source

Url : http://cve.mitre.org/cgi-bin/cvename.cgi?name=CVE-2023-25931

CWE : Common Weakness Enumeration

% Id Name
100 % CWE-287 Improper Authentication

CPE : Common Platform Enumeration

TypeDescriptionCount
Application 1
Application 1

Sources (Detail)

Source Url
MISC https://global.medtronic.com/xg-en/product-security/security-bulletins/pelvic...

Alert History

If you want to see full details history, please login or register.
0
1
2
Date Informations
2023-11-07 21:29:56
  • Multiple Updates
2023-03-09 05:27:17
  • Multiple Updates
2023-03-02 00:27:15
  • First insertion